Total Case Number: 10
Case 1: "No feed water" malfunction can be caused by a blocked conductivity water filter.
Case 2: "No feed water" malfunction can be caused by too much bio-film in the clean water tank.
Case 3: This is why the drainage has been blocked.
Case 4: For the premium model, do not put pouch strip in front of autoclave which may affect the efficiency of vacuum pump and cause vacumm system fail.
Case 5: Heavy dust may affect the efficiency of the vacuum pump which may cause a vacuum system fail. That is why you autoclave need an annual service.
Before service
After service
Case 6: A blocked HEPA filter may cause drying or vent problem.
Case 7: Bought a new MELAG or Sirona autoclave, but feel the printer is too expensive. There is a way to save you $800 if you have an old Siltex printer. Contact us.
Case 8: MELAG / Sirona Premium door stuck after a completed cycle. One of solenoid valve needs to be replaced.
Case 9: An old door gasket may cause the vacuum test failed or dripping water from the door during cycles.
Case 10: "bad water quality" means your RO water filters need to be replaced if your autoclave is fed by RO system. The following video clip illustrates the procedure of replacing water filters.